03.19.2020 - By Scott D. Weingart, MD FCCM
"Triple C" Circuit Disconnects
May be even higher risk than intubation if you do it wrong
Confer
Preplan with your partner
Clamp
Either non-marring clamp or clamp around 4x4 or tape on ETT (if disconnecting proximal to viral filter, than you can ignore the clamp)
Cut Flow
Put the vent in standby or disconnect the inspiratory limb
EMCrit CPAP Set-Up without the NC
See the Video
Here is a Pict
Here is a Pict with the 22mm OD Male-to-Male Connector and Corrugated Tubing
Better way to use the 3-way Stopcock
I showed in the video that you can use O2 tubing with 3-way stopcock but there is a clean oxygen leak. Bill Murphy, medic extraordinaire, wrote me with a better way. Cut off the fat part of the O2 line. You need to test with your hospital's lines to find out the best place to cut, but once you get rid of protrusion, you can make a tight connection without the multipurpose tube to luer adaptor.
Video Proving that BVM alone doesn't supply enough forward flow
and that even on flush-rate doesn't provide dangerous flow to patient to cause aerosol
Cardiac Arrest
Your safety then their safety
Safer Airway Society Guidelines
By all my SMACC airway buddies
Prepress Version
There is also a link to the article as well as printable infographics at the SaferAirways Site
Now on to the Podcast...