Dr. Howard Smith Reports

True METRIX Air Blood Glucose Meter Recalled


Listen Later

 

Vidcast:  https://youtu.be/No2ULvUr33k

 

Trividia Health, Inc. has a US nationwide recall on its TRUE METRIX® AIR Blood Glucose Meter.  One of the meters included in the True METRIX Blood Glucose Meter kit has an incorrect unit of measure as a factory default.  The unit measures glucose as mmol/L rather than the standard mg/dl.  If incorrectly read, a consumer will assume  a falsely low glucose level.  This could lead to incorrect glucose dosing.  If you have this device, check the serial number on the back of the unit.  The affected serial number is TA1548753.  If you have this meter, stop using it and call the company at 1-800-518-5726 to verify the serial number and to arrange a replacement.

 

https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts/trividia-health-inc-issues-nationwide-voluntary-recall-isolated-true-metrix-air-blood-glucose-meter#recall-announcement

 

#recall #glucosemeter #truemetrix #diabetes #insulin

 